Doctoral dissertation

Influence of patient shielding on dose in selected projections during conventional radiography

Author(s): Nejc Mekiš (Author), Dejan Žontar (Supervisor)

Thesis defense date: 23.09.2015

Organization: MPŠ - Mednarodna podiplomska šola Jožefa Stefana

PID: 20.500.12556/ReVIS-15070

Views: 6 | Downloads: 0

Abstract

Introduction: Due to the wide variety of use of x-rays in medicine, it is nowadays difficult to imagine medicine without them. In addition to its beneficial effects in diagnostics and treatment, ionising radiation can also be harmful, even at a very low dose. Therefore, we need to ensure the lowest possible exposure of patients. In addition to other measures this can be achieved, also by the use of protective lead shielding. Based on the latest findings of ICRP (International Commission of Radiation Protection, 2007), the most radiosensitive organs are colon, lungs, bone marrow, stomach and breasts, with the tissue weighting factor of 0.12.
Purpose: The purpose of this study is to determine, by a systematic analysis, whether and to what extent, the use of protective lead shielding in conventional x-ray imaging of pelvis, cervical spine, head, thoracic organs, hand and knee, as well as in mammography, reduces the absorbed dose of ionising radiation on the surface of the skin near the radiosensitive organs, i.e. breasts, testicles, lungs and colon.
Methods: Descriptive and experimental methods of research were used. The research was conducted in three parts. In the first part, we evaluated the electronic dosimeter Unfors EDD 30 as the measuring equipment of choice. Next, we measured the absorbed dose of ionising radiation in radiosensitive organs (lungs, colon, breasts and testicles) on a whole-body phantom. The x-ray examinations were selected with regard to the frequency of their performance in Slovenia (thoracic x-ray imaging and mammography) and on the basis of the effective dose received by the patients in certain examinations (head, cervical spine and pelvic imaging). For a wider range of examinations and for a better comparison of the effect of the use of lead shielding in different tests, we also included two examinations with lower dose exposure, i.e. x-rays of hand and knee. Both, the first and the second part of the research were conducted at the Radiological laboratory of the Faculty of Health Sciences in Ljubljana. In the last part of the research we measured the absorbed dose on the surface of the same radiosensitive organs as in the laboratory conditions, but on patients that were referred to x-ray imaging to the Radiology Department of the Community Health Centre Ljubljana (unit “Center”). In this part the results from the second part of the research were verified in real clinical situations.
Results and discussion: In pelvic imaging, the use of lead shielding led to a significant (several µGy) reduction of the dose to the lungs (about 60%) and breasts (about 80%). However, the reduction of the dose to the colon was not statistically significant, indicating that the use of colon protection in pelvic imaging is ineffective. In thoracic imaging, the use of lead shielding proved to be effective in colon protection, where the received dose was reduced by several µGy (about 40%). In relative terms shielding of testicles was quite effective (at least 50% reduction) while in absolute terms the received dose was quite low even without the use of lead shielding (up to a few tenths of µGy). The use of lead shielding in cervical spine imaging led to a significant (a few µGy) dose reduction to lungs (at least 65%) and breasts (at least 70%), whereas on colon (80%) and testicles (at least 90%) the dose was reduced by only a few tenths up to one µGy. The use of lead shielding in hand and knee imaging in most cases caused the dose to be reduced below the detection limit, but the absolute dose values were low even without the use of lead shielding (up to a few tenths of µGy). In head imaging it was unfortunately not possible to perform measurements on patients, but the measurements on the phantom showed high efficiency of lead shielding, with the dose on all the organs included in the study (lungs, breasts, colon and testicles) were reduced to about one tenth. Considering the absolute values lead shielding is particularly important in lung protection, where the reduction can reach several 10 µGy. In mammographic imaging, the measurements on the phantom indicate importance of lung and colon protection in CC projection with magnification. In those cases lead shielding proved to be very effective (dose reduction by several 10 times), while the exposure of the two organs without the use of lead shielding was significant (several µGy).
Conclusion: According to the results of the absolute and relative reductions of dose received by the selected radiosensitive organs, the conclusion can be made that the use of lead shielding is necessary in examinations with a “high dose level” for patients, which in this study included the imaging of pelvis, cervical spine and head. In the examinations with the so-called “medium dose level” (in my research defined as those with the dose to radiosensitive organs between about 1 and 10 µGy) to the selected radiosensitive organs, i.e. breasts, lungs and colon, in the imaging of thorax, cerebral spine and in mammography with magnification, it is recommended to use lead shielding, since the already low dose of ionising radiation is further reduced, often below the dosimeter detection limit. In the examinations with the so-called “low dose” (hand and knee imaging), the nature of protection of the selected radiosensitive organs is primarily psychological. Although the relative reduction is often great, the absolute value of reduction is only a few tenths or even hundredths of µGy.